TDM Logo

HOBO 250-Foot Depth Water Level Data Logger

Overview

Menawarkan akurasi tinggi dengan harga terjangkau serta kemudahan penggunaan, tanpa memerlukan tabung ventilasi atau desikan yang rumit untuk perawatan, data logger ini ideal untuk merekam ketinggian air dan suhu hingga kedalaman 250 kaki pada sumur dangkal, sungai, danau, serta lahan basah air tawar (stainless steel), serta lahan basah air asin/payau dan area pasang surut (titanium).

Features

  • Perlindungan terhadap petir
  • Perangkat lunak HOBOware Pro memudahkan konversi data menjadi pembacaan ketinggian air yang akurat
  • Sampling multi-rate
  • Tersedia dalam empat pilihan rentang kedalaman
  • Sensor tekanan keramik yang tahan lama
  • Termasuk sertifikat kalibrasi 3-point NIST-traceable

Specifications

Pressure and Water Level Measurements U20-001-03 and U20-001-03-Ti

Operation Range 0 to 850 kPa (0 to 123.3 psia); approximately 0 to 76.5 m (0 to 251 ft) of water depth at sea level, or 0 to 79.5 m (0 to 262 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range 69 to 850 kPa (10 to 123.3 psia), 0° to 40°C (32° to 104°F)
Burst Pressure 1200 kPa (174 psia) or 112 m (368 ft) depth
Water Level Accuracy*
Typical error: ±0.05% FS, 3.8 cm (0.125 ft) water
Maximum error: ±0.1% FS, 7.6 cm (0.25 ft) water
Raw Pressure Accuracy** ±0.3% FS, 2.55 kPa (0.37 psi) maximum error
Resolution <0.085 kPa (0.012 psi), 0.87 cm (0.028 ft) water
Pressure Response Time (90%)*** <1 second; measurement accuracy also depends on temperature response time

Temperature Measurements (All Models)

Operation Range -20° to 50°C (-4° to 122°F)
Accuracy ±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F), see Plot A in manual
Resolution 0.10°C at 25°C (0.18°F at 77°F), see Plot A in manual
Response Time (90%) 5 minutes in water (typical)
Stability (Drift) 0.1°C (0.18°F) per year

Logger

Real-time Clock
± 1 minute per month 0° to 50°C (32° to 122°F)
Battery 2/3 AA, 3.6 Volt lithium, factory-replaceable
Battery Life (Typical Use) 5 years with 1 minute or greater logging interval
Memory (Non-volatile) 64K bytes memory (approx. 21,700 pressure and temperature samples)
Weight
Stainless steel models: approximately 210 g (7.4 oz)
Titanium models: approximately 140 g (4.8 oz)
Dimensions 2.46 cm (0.97 inches) diameter, 15 cm (5.9 inches) length; mounting hole 6.3 mm (0.25 inches) diameter
Wetted Materials 316 stainless steel, Viton® o-rings, acetyl cap, ceramic sensor
Logging Interval Fixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from 1 second to 18 hours. Refer to the HOBOware software manual.
Launch Modes Immediate start and delayed start
Offload Modes Offload while logging; stop and offload
Battery Indication Battery voltage can be viewed in status screen and optionally logged in datafile. Low battery indication in datafile.
Environmental Rating IP68
* Water Level Accuracy: With accurate reference water level measurement, known water density, accurate Barometric Compensation Assistant data, and a stable temperature environment.
** Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors.
*** Changes in Temperature: Allow 10 minutes in water to achieve full temperature compensation of the pressure sensor. Maximum error due to rapid thermal changes is approximately 0.5%.