Start typing to see products you are looking for.

Alat Uji Tarik dan Tekan Elektronik Universal WDW-10E Microcomputer Control Electronic Universal Testing Machine

No. Technical indicators WDW-10E WDW-20E
1 Testing machine grade Grade 1 or 0.5
2 Maximum test force 10kN 20 kN
3 Test force measurement range Maximum test force ( 0.4% ~ 100% )
4 Test force measurement accuracy Within the measurement range of the maximum test force ( 0.4% to 100% ), it is better than ±1 % of the indicated value ( level 1 );

Within the measurement range of the maximum test force ( 0.4% to 100% ), it is better than ±0.5 % of the indicated value ( 0.5 grade)

5 Test power resolution ±300000 yards
6 deformation maximum 10mm ; large deformation extensometer 800mm
7 Deformation measurement range 0.2%-100% of the full scale of the extensometer
8 Deformation measurement accuracy Within the range of 0.2%-100% of the full scale of the extensometer , it is better than ±1 % of the indicated value (Class 1 );

Within the range of 0.2%-100% of the full scale of the extensometer , it is better than ±0.5 % of the indicated value (class 0.5 )

9 Displacement measurement accuracy Better than ±1 % of the indicated value ( grade 1 ); better than ±0.5 % of the indicated value ( grade 0.5 )
10 Displacement resolution 0.001mm
11 Displacement speed (mm/min) 0. 005 ~ 500 , stepless speed regulation
12 Stretching stroke (mm) 800
13 Compression stroke (mm) 800
14 Test width (mm) 370
15 Force control speed adjustment range 0.005–5%Fs/s
16 Force control rate control accuracy Within ± 1% of the set value
17 Deformation rate adjustment range 0.005—5%Fs/s
18 Deformation rate control accuracy When the rate is less than 0.05%Fs , it is within ±1% of the set value;

When the rate is greater than 0.05%Fs , it is within ±0.5% of the set value

19 Displacement rate control accuracy When the speed is less than 0.01mm/min , it is within ±1% of the set value ;

When the speed is  0.01mm/min , it is within ±0.5% of the set value .

20 Constant force, constant displacement, constant deformation control range 0.5%~100%Fs
21 Constant force, constant displacement, constant deformation control accuracy When the set value is greater than or equal to 10%Fs , it is within ±0.1% of the set value ;

When the rate is less than 10%Fs , within ±1% of the set value

22 working environment Room temperature 10 ~ 35℃ , humidity 20% ~ 80%
23 Mainframe Dimensions

(W×D×H) (mm)

686 × 585 × 1880
24 Host weight 250 250