TDM Logo

Coating Thickness Gauge Probes

* Harga merupakan harga standar, harga dapat berubah, silakan tanyakan terlebih dahulu dengan hubungi kami di bawah

Probe Opsional dan Panduan Spesifikasi untuk Pengukur Ketebalan Lapisan.

Optional Probes and Specification Guide for Coating Thickness Gauge

Probe model F400 F1 F1/90Âş F10 N1 CN02
Operating principle Magnetic induction Eddy current
Measuring range (μm) 0-400 0-1250  0-10000 0 to 1250 μm
0 to 40ÎĽm
(for chrome plate
on copper)
10~200
Low range resolution (ÎĽm) 0.1 0.1 10 0.1 1
Accuracy One-point calibration (μm) ±(3%H+1) ±(3%H+10) ±(3%H+1.5) ±(3%H+1)
Two-point calibration (μm) ±[(1~3)H%+0.7] ±[(1~3)H%+1] ±[(1~3)%H+10] ±[(1~3)%H+1.5]
Measuring conditions Min curvature of the
min area (mm)
Convex 1 1.5 Flatten 10 3 Flatten
Diameter of the min
area (mm)
φ3 φ7 φ7 φ40 φ5 φ7
Critical thickness
of substrate (mm)
0.2 0.5 0.5 2 0.3 unlimited

Reference Table for Probe selection

                                     Coating

Substrate

Non-magnetism coatings
(Organic Materials)
Non-magnetism coatings (non ferrous metals)
Thickness of coating is more than 100 ÎĽm Thickness of coating is less than 100 ÎĽm Thickness of coating is more than 100 ÎĽm Thickness of coating is less than 100 ÎĽm
Steel, iron and other magnetism metal Diameter of testing area is more than 30 mm F1 Probe: 0-1250 ÎĽm

f-400 probe: 0-400 ÎĽm

F1 Probe: 0-1250 ÎĽm

F10 probe: 0-10 ÎĽm

F1 Probe: 0-1250 ÎĽm

f-400 probe: 0-400 ÎĽm

F1 Probe: 0-1250 ÎĽm

F10 probe: 0-10 ÎĽm

Diameter of testing area is less than 30 mm f-400 probe: 0-400 ÎĽm F1 Probe: 0-1250 ÎĽm

f-400 probe: 0-400 ÎĽm

f-400 probe: 0-400 ÎĽm F1 Probe: 0-1250 ÎĽm

f-400 probe: 0-400 ÎĽm

Copper, Brass, Alumunium, Zinc, Tin and other metallic Diameter of testing is more than 5 mm N1 probe: 0-1250 ÎĽm N1 probe: 0-40 ÎĽm (for chrome plate on chopper)
Plastic non metallic substrate Diameter of testing is more than 70 mm CN 02 Probe: 10-200 ÎĽm