* Harga merupakan harga standar, harga dapat berubah, silakan tanyakan terlebih dahulu dengan hubungi kami di bawah
Probe Opsional dan Panduan Spesifikasi untuk Pengukur Ketebalan Lapisan.
Optional Probes and Specification Guide for Coating Thickness Gauge
| Probe model | F400 | F1 | F1/90º | F10 | N1 | CN02 | |||||||
| Operating principle | Magnetic induction | Eddy current | |||||||||||
| Measuring range (μm) | 0-400 | 0-1250 | 0-10000 | 0 to 1250 μm 0 to 40μm (for chrome plate on copper) |
10~200 | ||||||||
| Low range resolution (μm) | 0.1 | 0.1 | 10 | 0.1 | 1 | ||||||||
| Accuracy | One-point calibration (μm) | ±(3%H+1) | ±(3%H+10) | ±(3%H+1.5) | ±(3%H+1) | ||||||||
| Two-point calibration (μm) | ±[(1~3)H%+0.7] | ±[(1~3)H%+1] | ±[(1~3)%H+10] | ±[(1~3)%H+1.5] | – | ||||||||
| Measuring conditions | Min curvature of the min area (mm) |
Convex 1 | 1.5 | Flatten | 10 | 3 | Flatten | ||||||
| Diameter of the min area (mm) |
φ3 | φ7 | φ7 | φ40 | φ5 | φ7 | |||||||
| Critical thickness of substrate (mm) |
0.2 | 0.5 | 0.5 | 2 | 0.3 | unlimited | |||||||
Reference Table for Probe selection
| Coating
Substrate |
Non-magnetism coatings (Organic Materials) |
Non-magnetism coatings (non ferrous metals) | |||
| Thickness of coating is more than 100 μm | Thickness of coating is less than 100 μm | Thickness of coating is more than 100 μm | Thickness of coating is less than 100 μm | ||
| Steel, iron and other magnetism metal | Diameter of testing area is more than 30 mm | F1 Probe: 0-1250 μm
f-400 probe: 0-400 μm |
F1 Probe: 0-1250 μm
F10 probe: 0-10 μm |
F1 Probe: 0-1250 μm
f-400 probe: 0-400 μm |
F1 Probe: 0-1250 μm
F10 probe: 0-10 μm |
| Diameter of testing area is less than 30 mm | f-400 probe: 0-400 μm | F1 Probe: 0-1250 μm
f-400 probe: 0-400 μm |
f-400 probe: 0-400 μm | F1 Probe: 0-1250 μm
f-400 probe: 0-400 μm |
|
| Copper, Brass, Alumunium, Zinc, Tin and other metallic | Diameter of testing is more than 5 mm | N1 probe: 0-1250 μm | N1 probe: 0-40 μm (for chrome plate on chopper) | ||
| Plastic non metallic substrate | Diameter of testing is more than 70 mm | – | – | CN 02 Probe: 10-200 μm | |